FDA Device recall Z-2500-2015
NovaPlus Trach Care* Closed Suction System for Adults, T-Piece, 14F (Product Code 2V220135)
- FDA Device
- Class II
- terminated
- Published 2015-09-09
On 2015-09-09 the FDA classified a Class II recall of NovaPlus Trach Care* Closed Suction System for Adults, T-Piece, 14F (Product Code 2V220135) by Halyard Health, citing the thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" posit….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2500-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-29 |
| Published | 2015-09-09 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Halyard Health |
| Distribution | US |
Product
NovaPlus Trach Care* Closed Suction System for Adults, T-Piece, 14F (Product Code 2V220135); NovaPlus Trach Care* Closed Suction System for Adults, Clear T-Piece, 14F (Product Code 2V2205) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.
Reason
The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.
Identifiers
| code_info | Product Code 2V220135 - M5081T602, M5089T607, M5089T628, M5106T601; Product Code 2V2205 - M5079T605, M5089T608 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2500-2015%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.