FDA Device recall Z-2499-2019
Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a co…
- FDA Device
- Class II
- terminated
- Published 2019-09-18
On 2019-09-18 the FDA classified a Class II recall of Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a co… by Exactech, citing the threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2499-2019 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2019-07-24 |
| Published | 2019-09-18 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Exactech |
| Distribution | AR, CA, FL, IL, IN, NY, SC, TX |
Product
Threaded Offset Cup Impactor - Product Usage: To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use.
Reason
The threaded mechanism of the Threaded Offset Cup Impactor may be gritty and difficult to turn potentially due to entrapped blast media.
Identifiers
| code_info | Catalog Number 181-02-01, Lot Numbers 103071001, UDI Number 10885862520623. Expansion on 09/09/2019 to include Lot Number 8182001 |
|---|