RecallRadar

FDA Device recall Z-2494-2015

KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A tracheobronchial suction catheter is a device…

On 2015-09-09 the FDA classified a Class II recall of KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A tracheobronchial suction catheter is a device… by Halyard Health, citing the thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" posit….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2494-2015
ClassificationClass II
Statusterminated
Initiated2015-07-29
Published2015-09-09
Last updated2026-09-13 11:53 UTC
CompanyHalyard Health
DistributionUS

Product

KimVent* Wet Pak* Closed Suction System for Adults, 14 F, DT, T-Piece (Product Code 220569) A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.

Reason

The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the "LOCKED" position.

Identifiers

code_infoProduct Code 220569 - AB5145U14, AB5166U09, AB5180U11

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2494-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.