FDA Device recall Z-2492-2026
MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
- FDA Device
- Class II
- ongoing
- Published 2026-06-24
On 2026-06-24 the FDA classified a Class II recall of MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839 by Medline Industries, citing during an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surfa….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2492-2026 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2026-05-04 |
| Published | 2026-06-24 |
| Last updated | 2026-09-13 11:58 UTC |
| Company | Medline Industries |
| Distribution | US |
Product
MEDLINE Medical Procedure Kits labeled as: CATH 22F ROBINSON STRL 5'S PK, Medline Kit Number/SKU DYNJ0040839
Reason
During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
Identifiers
| code_info | UDI/DI 10889942046126 (each), 40889942046127 (case), Lot Numbers: 25JMK709, 25FMJ661, 25CMG042, 24IME549, 24GMD004, 24CMF091, 24CME465, 24BMH490. |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2492-2026%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.