FDA Device recall Z-2490-2015

7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY

On 2015-09-09 the FDA classified a Class II recall of 7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY by Trilliant Surgical, citing the affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2490-2015
ClassificationClass II
Statusterminated
Initiated2015-08-06
Published2015-09-09
Last updated2026-09-13 11:53 UTC
CompanyTrilliant Surgical
DistributionCA, FL, MN, NJ, NM, OH, WI

Product

7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.

Reason

The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates. The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.

Identifiers

code_infoLots: TSL002595, TSL002596, TSL002597