RecallRadar

FDA Device recall Z-2487-2012

Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland

On 2012-10-10 the FDA classified a Class II recall of Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland by Stryker Howmedica Osteonics, citing stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2487-2012
ClassificationClass II
Statusterminated
Initiated2012-04-02
Published2012-10-10
Last updated2026-09-13 11:53 UTC
CompanyStryker Howmedica Osteonics
DistributionUS

Product

Stryker Orthopaedics Restoration Modular Cone Body Trials Stryker Ireland Carrigtwohill Industrial Estate Carrigtwohill County Cork, Ireland. The Cone Body Trials simulate the respective Cone Body implants and mate with the stem trials to assist the surgeon in assessing the fit of the proposed Implants. The cone body trials is an assembly of the Body and 8 mm Hex Locking Bolt. The trial stem is attached to the trial Cone body via the distal thread of the Locking bolt.

Reason

Stryker Orthopaedics has become aware of a burr, or a metal shaving, observed in the internal threaded portion of the Cone Body Trial. The typical size of the shaving observed was 1-2 mm in length and 0.037 mm squared in cross section.

Identifiers

code_info
Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm…Catalog No. 6278-1-019 Mod Rev 19 mm std cone trial 6278-1-021 Mod Rev 21 mm std cone trial 6278-1-023 Mod Rev 23 mm std cone trial 6278-1-025 Mod Rev 25 mm std cone trial 6278-1-027 Mod Rev 27 mm std cone trial 6278-1-029 Mod Rev 29 mm std cone trial 6278-1-031 Mod Rev 31 mm std cone trial 6278-1-119 Mod Rev 19 mm + 10 mm cone trial 6278-1-121 Mod Rev 21mm + 10 mm cone trial 6278-1-123 Mod Rev 23mm + 10 mm cone trial 6278-1 -125 Mod Rev 25 mm + 10 mm cone trial 6278-1-127 Mod Rev 27mm + 10 mm cone trial 6278-1-129 Mod Rev 29mm + 10mm cone trial 6278-1-131 Mod Rev 31mm + 10 mm cone trial 6278 -1-219 Mod Rev 19mm + 20 mm cone trial 6278-1-221 Mod Rev 21mm +20 mm cone trial 6278-1-223 Mod Rev 23mm + 20 mm cone trial 6278-1-225 Mod Rev 25mm + 20 mm cone trial 6278-1-227 Mod Rev 27mm + 20 mm cone trial 6278-1-229 Mod Rev 29mm + 20 mm cone trial 6278-1-231 Mod Rev 31mm + 20 mm cone trial 6278-1-319 Mod Rev 19mm + 30 mm cone trial 6278-1-321 Mod Rev 21mm +30 mm cone tr

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2487-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.