RecallRadar

FDA Device recall Z-2486-2012

Optilock T15 AO Driver (3.5mm)

On 2012-10-10 the FDA classified a Class II recall of Optilock T15 AO Driver (3.5mm) by Biomet, citing field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2486-2012
ClassificationClass II
Statusterminated
Initiated2012-08-23
Published2012-10-10
Last updated2026-09-13 11:53 UTC
CompanyBiomet
DistributionUS

Product

Optilock T15 AO Driver (3.5mm). Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.

Reason

Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw

Identifiers

code_infoCatalog number: 36505; and lot numbers: 069540, 179190, 201610, 385380, 419940, 427530, 440360, 440370, 555370, 638560, 645820, and 956960.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2486-2012%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.