FDA Device recall Z-2485-2012
Optilock 2.7 mm Screw Inserter
- FDA Device
- Class II
- terminated
- Published 2012-10-10
On 2012-10-10 the FDA classified a Class II recall of Optilock 2.7 mm Screw Inserter by Biomet, citing field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2485-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-08-23 |
| Published | 2012-10-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Biomet |
| Distribution | US |
Product
Optilock 2.7 mm Screw Inserter. Intended Use: The OptiLock Upper Extremity Plating System is intended for fixation of fractures and osteotomies. The System is intended for fractures and fracture dislocations, osteotomies and non-unions of the proximal humerus, particularly in osteopenic bone. The System is intended for fixation of fractures, osteotomies and non-unions of the olecranon, humerus, radius, ulna, particularly in osteopenic bone. The OptiLock VL Distal Radius Plating System is intended for use in distal radius and ulna (and other small bones), including use in osteopenic bone.
Reason
Field complaints indicate that recent shipments of the 14-400795 and 36505 drivers do not properly mate ("lock") into some of the specified screw heads. Drivers near MMC (Maximum Material Condition) or that are misshapen may not properly mate with screws at MMC. Design dimensions and tolerances of the drivers are not properly specified to prevent an incorrect fit between the drivers and the screw
Identifiers
| code_info | Catalog number: 14-400795; and lot numbers: 133420, 409460, 469960, 487650, 487660, 627940, 627950, 797190, and 797200. |
|---|