FDA Device recall Z-2484-2015

OmniPod¿ Insulin Management System

On 2015-09-16 the FDA classified a Class I recall of OmniPod¿ Insulin Management System by Insulet, citing omniPods¿ (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2484-2015
ClassificationClass I
Statusterminated
Initiated2015-07-13
Published2015-09-16
Last updated2026-09-13 11:53 UTC
CompanyInsulet
DistributionUS

Product

OmniPod¿ Insulin Management System. Only the OmniPod¿ device is affected. Catalog Number(s): POD-ZXP420 (US). Intended for subcutaneous delivery of insulin at set and variable rates for the management of diabetes mellitus in persons requiring insulin.

Reason

OmniPods¿ (Pods) have a higher rate of failure causing: Cannula fails to deploy/retract causing insulin to be pumped or the audible alarm is displayed on the PDM and pod will not deliver insulin

Identifiers

code_info
Lot Numbers: L40806, L40811, L40895, L40976, L41014, L41025, L41067, L41162, L41171, L41197, L41198, L41250 Expa…Lot Numbers: L40806, L40811, L40895, L40976, L41014, L41025, L41067, L41162, L41171, L41197, L41198, L41250 Expanded Recall: Lot Codes: L41880, L41881, L41892, L41895,L41897,L41898, L41899,L41900, L41901, L41902, L41903, L41904, L41905, L41906, L41907

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.