FDA Device recall Z-2480-2012
00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS…
- FDA Device
- Class II
- terminated
- Published 2012-10-10
On 2012-10-10 the FDA classified a Class II recall of 00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS… by Zimmer, citing complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in s….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2480-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2011-04-01 |
| Published | 2012-10-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Zimmer |
| Distribution | US |
Product
00780302601 MIS 12/14 Slotted Femoral Head Provisional -3.5X26 00780302602 MIS 12/14 Slotted Femoral Head Provisional +0X26 00780302603 MIS 12/14 Slotted Femoral Head Provisional +3.5X26 00780302604 MIS 12/14 Slotted Femoral Head Provisional +7X26 00780302605 MIS 12/14 Slotted Femoral Head Provisional +10.5X26 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
Reason
Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
Identifiers
| code_info | All lots including lot 78939000 (1/7/2002) to present |
|---|