Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological…
- FDA Device
- Class II
- terminated
- Published 2018-07-25
On 2018-07-25 the FDA classified a Class II recall of Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological… by Hitachi Medical Systems America, citing the stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received tw….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2479-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-18 |
| Published | 2018-07-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Hitachi Medical Systems America |
| Distribution | US |
Product
Hitachi Oasis MRI system Product Usage: Hitachi MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved crosssectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Reason
The stainless steel belt which runs the length of the table underneath the tabletop is exposed when the tabletop is extended into the magnet for scanning. The firm has received two complaints where technologists have cut their fingers on this belt.
Identifiers
| code_info | M001 to M028, M030 to M108, M110 to M266, M951, M952 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2479-2018%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.