FDA Device recall Z-2472-2021

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use

On 2021-09-22 the FDA classified a Class II recall of Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use by Abbott Laboratories, citing due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2472-2021
ClassificationClass II
Statusongoing
Initiated2021-07-26
Published2021-09-22
Last updated2026-09-13 11:56 UTC
CompanyAbbott Laboratories
DistributionUS

Product

Alinity ci-series System Control Module clinical chemistry and immunoassay analyzer for in vitro diagnostic use. Model number LN 03R70-01

Reason

Due to potential performance issues with software version 3.2.3 and earlier. Performance issues are: 1) There is the potential for onboard reagents past their lot expiration date or onboard stability time to be incorrectly displayed on the Reagent Status screen with a status of "OK". In this scenario, the expired reagent will remain in the carousel for sample processing. 2) Performing monthly maintenance procedure 5701 Clean ICT Drain Tip incorrectly may potentially cause issues such as damaged connector or leaking connection. 3) A sample exception with message code 150 "Unable to process test. Previous processing module error" is produced without notifying the operator of a reagent pipettor failure and may cause processing module to transition into the Pausing state. Tests initiated prior to the sample exception may continue to process without dispense of reagents, potentially leading to incorrect results. 4. Reagent short dispense: Alinity c needs a control measure to protect against the potential for short reagent dispense. Addition of a maximum depth limit for reagent cartridges on the Alinity c to prevent the potential for a short dispense is needed. 5. RSM indicators were blinking green incorrectly: The Reagent and Sample Manager (RSM) indicators will incorrectly blink green indicating that processing is completed and the rack or cartridge can be removed from the loading area when the rack or cartridge should not be accessed. 6. Erroneous ICT Calibration Curves: Erroneous ICT calibration curves might be generated for the Alinity c when there is a malfunction of ICT reference solution delivery, use of wrong/evaporated calibrator, defective ICT modules, etc.

Identifiers

code_info
Model Number: LN 03R70-01 all units with software versions 3.2.3 and earlier. Serial Numbers manufactured by Sanmin…Model Number: LN 03R70-01 all units with software versions 3.2.3 and earlier. Serial Numbers manufactured by Sanmina: SCM01001; SCM01002; SCM01003; SCM01004; SCM01005; SCM01006; SCM01007; SCM01008; SCM01009; SCM01010; SCM01011; SCM01012; SCM01013; SCM01014; SCM01015; SCM01016; SCM01017; SCM01018; SCM01019; SCM01020; SCM01021; SCM01022; SCM01023; SCM01024; SCM01025; SCM01026; SCM01027; SCM01028; SCM01029; SCM01030; SCM01031; SCM01032; SCM01033; SCM01034; SCM01035; SCM01036; SCM01037; SCM01038; SCM01039; SCM01040; SCM01041; SCM01042; SCM01043; SCM01044; SCM01045; SCM01046; SCM01047; SCM01048; SCM01049; SCM01050; SCM01051; SCM01052; SCM01053; SCM01054; SCM01055; SCM01056; SCM01057; SCM01058; SCM01059; SCM01060; SCM01061; SCM01062; SCM01063; SCM01064; SCM01065; SCM01066; SCM01067; SCM01068; SCM01069; SCM01070; SCM01071; SCM01072; SCM01073; SCM01074; SCM01075; SCM01076; SCM01077; SCM01078; SCM01079; SCM01080; SCM01081; SCM01082; SCM01083; SCM01084; SCM01085; SCM01086; SCM01087; SCM01088

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.