FDA Device recall Z-2471-2024
1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1
- FDA Device
- Class II
- ongoing
- Published 2024-08-14
On 2024-08-14 the FDA classified a Class II recall of 1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1 by Ge Medical Systems, citing GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the fol….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2471-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-06-14 |
| Published | 2024-08-14 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Ge Medical Systems |
| Distribution | n/a |
Product
1.5T SIGNA HDxt Family, with affected software versions: HD16.0_V03 to HD16.4, HD23.0_V03, HD28.0, HD29.1, MR30.0, MR30.1; Nuclear Magnetic Resonance Imaging System
Reason
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
Identifiers
| code_info | System ID: 907373VMR, 256SPMR2, 870642MR, 5340995MR11, KRMCMR2, SMSCIA2, 714229MR2, SHIMR02637, 310358VMR, RADNETMR664…System ID: 907373VMR, 256SPMR2, 870642MR, 5340995MR11, KRMCMR2, SMSCIA2, 714229MR2, SHIMR02637, 310358VMR, RADNETMR664, 559447MRVOYA3, 559447MRVOYA2, 559447MRVOYA1, INSIGHTMR1397, 626359VMR4, INSIGHTMR1371, ALLIANCEVOYA04, 707431MR, 323294VMR, 310823VMR, SHIMR02636, 818496HCMR1, 209SUTBMR, SHIMR02625, SHIMR02617, 530OROMMR, SHIMR02645, INSIGHTMR1385, INSIGHTMR1378, SHIMR02647, 13412, SHIMR02648, SHIMR02635, SHIMR02654, RADNETMR648, SHIMR02639, ALLIANCEVOYA11, INSIGHTMR1376, SHIMR02650, SHIMR02638, SHIMR02624, 818708MR2, 707462VOYMR, 661HMVMR1, 760247VMR, SHIMR02653, 303750MR, 307630MR2, 719394MR, 970242VAMR, 3036140MR, 720513MR, 970879VMR, 860289VOYAGER, 860659MRVOYAGER, 203453VOYAGER, 860347VOYAGER, 203834VOYAGER, 302655MR1, 561996MR1, 813315MR, 352597MR1, ALLIANCEVOYA01, 305445MR1, 904CSMR, 904627MR3, SHIMR02652, 321CLMRVOY, 305229MR1, 239NHVOY, 352ADVMR, 352867MR2, 941MPGPGVOY, 813655MR, 813874MR1, 941VOYMR, 850325MR1, 850877MR1, 229758MRI, 770539MR, MSF138-1-6XIAHPU7, 478745VMR, 2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2471-2024%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.