Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving and communications s…
- FDA Device
- Class II
- terminated
- Published 2015-09-02
On 2015-09-02 the FDA classified a Class II recall of Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving and communications s… by Ge Healthcare, citing images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2470-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-08-11 |
| Published | 2015-09-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Ge Healthcare |
| Distribution | US |
Product
Centricity Universal Viewer Product Usage: The Centricity Universal Viewer Version 6.0 software is a picture archiving and communications system, a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. Its hardware components may include workstations, digitizers, communications devices, computers, video monitors, magnetic, optical disk, or other digital data storage devices, and hardcopy devices. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. Used in Radiology.
Reason
Images from the Centricity PACS-IW with Universal Viewer and Centricity Universal Viewer may be missing when a system parameter MapRoute is set to a value greater than 1.
Identifiers
| code_info | Model Number: Centricity Universal Viewer Version 6.0 software Codes: 2090255-001 UV 6.0 Web Client Media Kit. 05 Jun 15 to present |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2470-2015%22