RecallRadar

FDA Device recall Z-2469-2025

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C

On 2025-09-03 the FDA classified a Class II recall of AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C by American Contract Systems, citing during an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for furthe….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2469-2025
ClassificationClass II
Statusongoing
Initiated2025-07-14
Published2025-09-03
Last updated2026-09-13 11:57 UTC
CompanyAmerican Contract Systems
DistributionIL, LA

Product

AMERICAN CONTRACT SYSTEMS, Medical Procedure kits and trays labeled as: 1) BORN ON ARRIVAL KIT, Model/Item Number LLBN11C; 2) LABOR & DELIVERY PACK, Model/Item Number LLLD19I.

Reason

During an internal investigation, ACS determined that the rationale for sterilization of certain components was not justified and therefore, some components are not fit for further sterilization according to ACS procedures. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

Identifiers

code_info
1) Model/Item Number LLBN11C: UDI/DI 00191072233530, Kit Lot Numbers: 70-051710, 70-052354; 2) Model/Item Number LLL…1) Model/Item Number LLBN11C: UDI/DI 00191072233530, Kit Lot Numbers: 70-051710, 70-052354; 2) Model/Item Number LLLD19I: UDI/DI 00191072233493, Kit Lot Numbers: 70-051324, 70-051967, 70-053006, 70-053303, 70-053710.

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2469-2025%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.