GS70 Salus Surgical Table, Product Code 18-070-70

On 2023-09-06 the FDA classified a Class II recall of GS70 Salus Surgical Table, Product Code 18-070-70 by Skytron, citing when the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2469-2023
ClassificationClass II
Statusongoing
Initiated2023-08-01
Published2023-09-06
Last updated2026-09-13 11:56 UTC
CompanySkytron
DistributionUS

Product

GS70 Salus Surgical Table, Product Code 18-070-70

Reason

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

Identifiers

code_info
Serial Numbers: SUC-20207-0006 SUC-20208-0025 SUC-20209-0044 SUC-20212-0063 SUC-20228-0082 SUC-2022Y-0101 SUC-20207-000…Serial Numbers: SUC-20207-0006 SUC-20208-0025 SUC-20209-0044 SUC-20212-0063 SUC-20228-0082 SUC-2022Y-0101 SUC-20207-0007 SUC-20208-0026 SUC-20209-0045 SUC-20212-0064 SUC-20228-0083 SUC-2022Y-0102 SUC-20207-0008 SUC-20208-0027 SUC-2020X-0046 SUC-20212-0065 SUC-20228-0084 SUC-2022Y-0103 SUC-20207-0009 SUC-20208-0028 SUC-2020X-0047 SUC-20226-0066 SUC-20228-0085 SUC-2022Y-0104 SUC-20207-0010 SUC-20208-0029 SUC-2020X-0048 SUC-20226-0067 SUC-20228-0086 SUC-2022Y-0105 SUC-20207-0011 SUC-20209-0030 SUC-2020X-0049 SUC-20226-0068 SUC-20228-0087 SUC-20231-0106 SUC-20207-0012 SUC-20209-0031 SUC-2020X-0050 SUC-20226-0069 SUC-20228-0088 SUC-20231-0107 SUC-20207-0013 SUC-20209-0032 SUC-2020X-0051 SUC-20226-0070 SUC-20228-0089 SUC-20231-0108 SUC-20208-0014 SUC-20209-0033 SUC-2020X-0052 SUC-20227-0071 SUC-20228-0090 SUC-20231-0109 SUC-20208-0015 SUC-20209-0034 SUC-2020X-0053 SUC-20227-0072 SUC-20229-0091 SUC-20231-0110 SUC-20208-0016 SUC-20209-0035 SUC-2020X-0054 SUC-20227-0073 SUC-20229-0092 SUC-20231