FDA Device recall Z-2468-2015
Stryker Sustainability Solutions CB5LT Ethicon Endo-Surgery, Inc
- FDA Device
- Class II
- terminated
- Published 2015-09-02
On 2015-09-02 the FDA classified a Class II recall of Stryker Sustainability Solutions CB5LT Ethicon Endo-Surgery, Inc by Stryker Sustainability Solutions, citing stryker Sustainability Solutions has received reports indicating that devices from one lot (lot# 3574704) were labeled as 100mm models (CB5LT) while the devices inside of the pack….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2468-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-07-10 |
| Published | 2015-09-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Stryker Sustainability Solutions |
| Distribution | US |
Product
Stryker Sustainability Solutions CB5LT Ethicon Endo-Surgery, Inc. Excel Trocar Sleeve w/Stability Sleeve 5mm x 100mm Reprocessed by Stryker Sustainability Solutions Product Usage: The Reprocessed Ethicon ENDOPATH¿ XCEL Universal Trocar Stability Sleeve is a sterile single patient use device with a radiolucent sleeve. The trocar sleeve contains two seals that accommodate instruments 5mm in diameter. Together, these two seals minimize gas leakage when instruments are inserted or withdrawn through the trocar. A stopcock valve is compatible with standard luer lock fitting and provides attachment for gas insufflation and desufflation. The stopcock is in the closed position when it is parallel to the sleeve.
Reason
Stryker Sustainability Solutions has received reports indicating that devices from one lot (lot# 3574704) were labeled as 100mm models (CB5LT) while the devices inside of the packaging were 75mm models (CB5ST).
Identifiers
| code_info | Lot number 3574704 UDI Code: (01)00885825014421(17)180310(10)3574704 Model Number: CB5ST (labeled as CB5LT) |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2468-2015%22