FDA Device recall Z-2467-2020
ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment…
- FDA Device
- Class II
- terminated
- Published 2020-07-08
On 2020-07-08 the FDA classified a Class II recall of ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment… by Radiometer Medical Aps, citing barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some ba….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2467-2020 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2020-05-11 |
| Published | 2020-07-08 |
| Last updated | 2026-09-13 11:55 UTC |
| Company | Radiometer Medical Aps |
| Distribution | US |
Product
ABL80-FLEX-393-839 Box Label, REF 393-839, IVD, CE, UDI: 05700693938394 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Reason
Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.
Identifiers
| code_info | All lots |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2467-2020%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.