RecallRadar

FDA Device recall Z-2455-2024

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination wit…

On 2024-08-07 the FDA classified a Class II recall of DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination wit… by Beckman Coulter Mishima K K, citing there are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same r….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2455-2024
ClassificationClass II
Statusongoing
Initiated2024-06-05
Published2024-08-07
Last updated2026-09-13 11:57 UTC
CompanyBeckman Coulter Mishima K K
DistributionUS

Product

DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Reason

There are two software bug issues with the analyzer. Issue 1: When a dedicated rack is reserved for calibration, after a calibration test has been placed and completed, the same rack may be used to place both a patient order and a calibration order. When a patient sample and a non-barcoded calibrator tube is processed on the rack, the calibration result will be reported as a patient result. Issue 2: calibration with expired calibrator. They both may cause erroneous patient results, but the probability of serious adverse health consequences or death is unlikely.

Identifiers

code_info
Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 20230800…Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers: 2023070010 2023060005 2023090015 2023070009 2023080011 2023080012 2023100019 2023100018 2023100020 2024010037 2024010036 2023110022 2024010038 2023090016 2024020040 2024020041 2024020039 2024020043 2024020042 2023110023 2024020045 2024020046 2024010024 2024010025

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2455-2024%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.