FDA Device recall Z-2453-2012
Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System…
- FDA Device
- Class II
- terminated
- Published 2012-10-03
On 2012-10-03 the FDA classified a Class II recall of Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System… by Hologic, citing fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM im….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2453-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-09-14 |
| Published | 2012-10-03 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Hologic |
| Distribution | US |
Product
Hologic Fluoroscan Mini C-Arm Model with InSight- FD Software Version 5.0.1and 5.0.2 The device is a Mini C-arm Fluoroscopic Imaging System designed to provide -physicians with general fluoroscopic visualization of a patient, including but not limited to surgical orthopedic and podiatry use, critical and emergency procedures, and light anatomy imaging situation.
Reason
Fluoroscan Mini C-Arm models -InSight2, version 4.0 and 4.0.1 - DICOM images stored in PACS may be flipped 180¿around vertical axis. InSight FD, version 5.0.1 and 5.0.2 - DICOM images stored in PACS may be lipped 180¿ around vertical axis as well as 180¿ around a horizontal axis.
Identifiers
| code_info | Software version 5.0.1and 5.0.2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2453-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.