FDA Device recall Z-2448-2015
LX3 Floor Stand for Endure Luxite Envision Optical System
- FDA Device
- Class II
- terminated
- Published 2015-09-02
On 2015-09-02 the FDA classified a Class II recall of LX3 Floor Stand for Endure Luxite Envision Optical System by Alcon Research, citing alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2448-2015 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2015-08-11 |
| Published | 2015-09-02 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Alcon Research |
| Distribution | US |
Product
LX3 Floor Stand for Endure Luxite Envision Optical System; Endure LuxOR Q-Vue. Product Usage: The LX3 Microscope Floor Stand is an accessory to the LuxOR Opthalmic Microscope and the LuxOR Opthalmic Microscope with Q-VUE 3-D Assistant Visualization and provides a portable suspension system for the microscope that integrates an AC power module and microscope controls (Primary Control Panel and connectivity of foot controllers)
Reason
Alcon is initiating a voluntary recall of all LX3 Microscope Floor Stands after receiving reports concerning loss of functionality due to issues with the cabling between the power, host and illumination modules, potentially causing the system to display a system message, and/or resulting in illumination failure.
Identifiers
| code_info | All units LX3 Microscope Floor Stands manufactured prior to July 2015 (Serial Numbers 1402548501X to 1502857606X) |
|---|