FDA Device recall Z-2446-2023
Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol,…
- FDA Device
- Class II
- ongoing
- Published 2023-08-30
On 2023-08-30 the FDA classified a Class II recall of Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol,… by Randox Laboratories, citing (1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2446-2023 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2023-06-21 |
| Published | 2023-08-30 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Randox Laboratories |
| Distribution | US |
Product
Randox Liquid Urine Control Level 2- intended for in vitro diagnostic use, in the quality control of Amylase, Calcium, Chloride, Cortisol, Creatinine, Glucose, hCG pregnancy, Magnesium, Microalbumin, Osmolality, pH, Phosphate Inorganic, Potassium, Protein Total, Sodium, Specific Gravity, Urea and Uric Acid (Urate) on clinical chemistry systems Catalog Number: UC5074
Reason
(1) There is vial to vial variation resulting in some vials recovering positive for hCG, which should be negative and (2) high and outside range for cortisol. (3) There has been a transcription error for Creatinine in the Instructions For Use (IFU). The target and ranges for Creatinine for the Roche Creatinine Plus method have been listed incorrectly. These failures cause a delay in patient results for the above mentioned analytes.
Identifiers
| code_info | GTIN: 05055273207569 Batch/Lot Number: 1209UC Expiry Date: 28 Mar 24 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2446-2023%22