FDA Device recall Z-2446-2014
CENTRAL LINE FULL BODY DRAPE CUSTOMED CONTENTS: (1) WRAPPER 30" X 30" (2) MASK SURGICAL W/ VISOR ANTI-FOG 160mm Hg (2) CAP NURSE 21" GREEN…
- FDA Device
- Class I
- terminated
- Published 2014-09-10
On 2014-09-10 the FDA classified a Class I recall of CENTRAL LINE FULL BODY DRAPE CUSTOMED CONTENTS: (1) WRAPPER 30" X 30" (2) MASK SURGICAL W/ VISOR ANTI-FOG 160mm Hg (2) CAP NURSE 21" GREEN… by Customed, citing customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to t….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2446-2014 |
| Classification | Class I |
| Status | terminated |
| Initiated | 2014-05-20 |
| Published | 2014-09-10 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Customed |
| Distribution | US |
Product
CENTRAL LINE FULL BODY DRAPE CUSTOMED CONTENTS: (1) WRAPPER 30" X 30" (2) MASK SURGICAL W/ VISOR ANTI-FOG 160mm Hg (2) CAP NURSE 21" GREEN BOUFFANT (2) TOWELS ABSORBENT 15" X 20" (2) GOWN LARGE STANDARD SMS VELCRO/NECK (1) DRAPE SHEET 41" X 58" SMS (1) DRAPE BRACHIAL ANGlO 44" X 31" REINFORCED 3" ADHESIVE CIRCULAR FENESTRATION Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Reason
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Identifiers
| code_info | Product code 900-2262,8 lots 112041231 113036345 113047112 113078310 113088797 140111207 140211823 140312128 |
|---|