FDA Device recall Z-2441-2018
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
- FDA Device
- Class II
- terminated
- Published 2018-07-25
On 2018-07-25 the FDA classified a Class II recall of Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope by Zeiss Carl, citing under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2441-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-05-29 |
| Published | 2018-07-25 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Zeiss Carl |
| Distribution | US |
Product
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Reason
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.