FDA Device recall Z-2441-2018

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

On 2018-07-25 the FDA classified a Class II recall of Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope by Zeiss Carl, citing under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2441-2018
ClassificationClass II
Statusterminated
Initiated2018-05-29
Published2018-07-25
Last updated2026-09-13 11:54 UTC
CompanyZeiss Carl
DistributionUS

Product

Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope

Reason

Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.

Identifiers

code_info432332-9020-000

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.