RecallRadar

FDA Device recall Z-2439-2020

Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993

On 2020-07-08 the FDA classified a Class II recall of Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993 by Livanova Deutschland, citing the SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossibl….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2439-2020
ClassificationClass II
Statusterminated
Initiated2019-08-26
Published2020-07-08
Last updated2026-09-13 11:55 UTC
CompanyLivanova Deutschland
DistributionFL, GA, IN, KY, MI, MO, NY, OH, PA, SC, TN, TX, VA

Product

Control Panel REF - 60-02-50: LivaNova SCP Pump Control Panel S5/C5. REF 60-02-50, IPX2, RxOnly, UDI: 04033817900993

Reason

The SCP control Panel rotary knob used to control rotary speed of the pump in the SCP system used as a centrifugal pump during cardiopulmonary bypass may be difficult or impossible to rotate as a result of premature wear of the angle encoder. The loss of function of the angle encoder during bypass may lead to ischemia, stroke, organ damage, arterial tissue injury or reperfusion syndrome.

Identifiers

code_infoSerial Number: 60S14735 60S14877 60S14715 60S14716 60S14736 60S14755 60S14707 60S14708

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2439-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.