FDA Device recall Z-2436-2023
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
- FDA Device
- Class I
- ongoing
- Published 2023-09-06
On 2023-09-06 the FDA classified a Class I recall of Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85 by Datascope, citing users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined th….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2436-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-07-31 |
| Published | 2023-09-06 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Datascope |
| Distribution | US |
Product
Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85
Reason
Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kinks/abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The firm's HHE also identified hardware issues related to these alarms.
Identifiers
| code_info | All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2436-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.