RecallRadar

FDA Device recall Z-2435-2023

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-…

On 2023-09-06 the FDA classified a Class I recall of Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-… by Datascope, citing users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined th….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2435-2023
ClassificationClass I
Statusongoing
Initiated2023-07-31
Published2023-09-06
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

Cardiosave Hybrid Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65

Reason

Users were reporting instances of "Gas Loss in IAB Circuit" and "Gas Gain in IAB Circuit" alarms while providing therapy. An internal investigation of the complaints determined that there is a potential trigger for these alarms that was not listed in the IFU: Patient Movement (coughing, general movement, and swallowing). Blood in the circuit, kins and abrasions of the tubing/catheter and certain patient conditions could also trigger these alarms. The HHE also identified hardware issues related to these alarms.

Identifiers

code_info
All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,…All Lot Numbers. Model: 0998-00-0800-31, UDI: 10607567109053; Model: 0998-UC-0800-31, UDI: N/A; Model: 0998-00-0800-32,UDI: 10607567111117; Model: 0998-00-0800-33, UDI: 10607567109008; Model: 0998-UC-0800-33, UDI: N/A; Model: 0998-00-0800-34, UDI: 10607567111940; Model: 0998-00-0800-35, UDI: 10607567109107; Model: 0998-00-0800-45, UDI: 10607567108421;Model: 0998-00-0800-52, UDI: 10607567108438; Model: 0998-UC-0800-52, UDI: N/A; Model: 0998-00-0800-53, UDI:10607567108391; Model: 0998-UC-0800-53, UDI: N/A; Model: 0998-00-0800-55, UDI: 10607567108414; Model: 0998-UC-0800-55,UDI: N/A; Model: 0998-00-0800-65, UDI: 10607567113432; Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2435-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.