Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask

On 2020-07-08 the FDA classified a Class II recall of Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask by Westmed, citing A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with i….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2435-2020
ClassificationClass II
Statusterminated
Initiated2019-01-15
Published2020-07-08
Last updated2026-09-13 11:55 UTC
CompanyWestmed
DistributionUS

Product

Neo-Vent Infant T-Piece Resuscitation Circuit With Infant Mask; Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0420, Qty:10, Length: 60" (1.5cm), Rx Only.

Reason

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Identifiers

code_info101618U02, 102318U06, 101518U03.