Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Pr…

On 2024-07-31 the FDA classified a Class II recall of Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Pr… by B Braun Medical, citing perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syri….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2434-2024
ClassificationClass II
Statusongoing
Initiated2024-06-14
Published2024-07-31
Last updated2026-09-13 11:57 UTC
CompanyB Braun Medical
DistributionUS

Product

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A

Reason

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Identifiers

code_infoLot Code: Model No 8713031U; UDI-DI 4046964797958; All Serial Numbers