RecallRadar

FDA Device recall Z-2434-2023

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-…

On 2023-09-06 the FDA classified a Class I recall of Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-… by Datascope, citing users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2434-2023
ClassificationClass I
Statusongoing
Initiated2023-07-31
Published2023-09-06
Last updated2026-09-13 11:56 UTC
CompanyDatascope
DistributionUS

Product

Cardiosave Rescue Intra-Aortic Balloon Pumps (IABP), Part Numbers 0998-00-0800-75, 0998-00-0800-83, 0998-00-0800-85, 0998-UC-0800-75, 0998-UC-0800-83, 0998-UC-0800-85,

Reason

Users reported "System Over Temperature" alarms associated with a loss of pumping and/or the Cardiosave system entering Standby mode.

Identifiers

code_info
All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-…All Lot Numbers. Model: 0998-00-0800-75, UDI: 10607567112312; Model: 0998-00-0800-83, UDI: 10607567108407; Model: 0998-00-0800-85, UDI: 10607567113449, Model: 0998-UC-0800-75, UDI: N/A; Model: 0998-UC-0800-83, UDI: N/A; Model: 0998-UC-0800-85, UDI: N/A;

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2434-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.