FDA Device recall Z-2431-2012
Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity i…
- FDA Device
- Class II
- terminated
- Published 2012-09-26
On 2012-09-26 the FDA classified a Class II recall of Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity i… by Sekisui Diagnostics, citing SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make mor….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2431-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-08-28 |
| Published | 2012-09-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Sekisui Diagnostics |
| Distribution | US |
Product
Sekisui Spectrolyse PAI-1, In-Vitro Diagnostic for Quantitative determination of Plasminogen Activator Inhibitor Type- 1 (PAI-1) activity in human plasma. Catalog Number: 101201 For the determination of Plasminogen Activator Inhibitor Type 1 (PAI-1) activity in human plasma. The test is for in Citro diagnostic use.
Reason
SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make more frequent sample dilutions.
Identifiers
| code_info | Lot Numbers: 120425 and 120601 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.