RecallRadar

FDA Device recall Z-2423-2018

Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable…

On 2018-07-18 the FDA classified a Class II recall of Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable… by Anika Therapeutics, citing hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the fi….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2423-2018
ClassificationClass II
Statusterminated
Initiated2018-05-10
Published2018-07-18
Last updated2026-09-13 11:54 UTC
CompanyAnika Therapeutics
DistributionUS

Product

Hyalomatrix, Product nos. 651105, 651103, 651104, and 651301 Product Usage: Hyalomatrix is a bi-layered, sterile, flexible, and comfortable wound dressing that acts as an advanced wound care device. It is comprised of a non-woven pad entirely composed of HY AFF 11, a benzyl ester of hyaluronic acid, and a semipermeable silicone membrane, which controls water vapor loss, provides a flexible covering for the wound surface, and adds increased tear strength to the device. The HYAFFl 1 wound contact layer biodegradable matrix acts as a scaffold for cellular invasion and capillary growth. Hyalomatrix is indicated for the management of wounds including: partial and full-thickness wounds; second-degree bums; pressure ulcers; venous ulcers; diabetic ulcers; chronic vascular ulcers; tunneled/undetermined wounds; surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence); trauma wounds (abrasions, lacerations, skin tears); and draining wounds. The device is intended for one-time use.

Reason

Hyalomatrix products are susceptible to degradation as a result of the decomposition of PVC in the packaging material during sterilization, which resulted in degradation of the film. Over the long term it is possible that product integrity may be compromised.

Identifiers

code_info2017017A, 2017033A, 2017034A, 2017031A, 2017065A, 2017066A, 2017007A, 2017008A, 2017030A

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2423-2018%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.