FDA Device recall Z-2420-2025
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2
- FDA Device
- Class II
- ongoing
- Published 2025-09-03
On 2025-09-03 the FDA classified a Class II recall of Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2 by Maquet, citing under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2420-2025 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2025-07-14 |
| Published | 2025-09-03 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | Maquet |
| Distribution | US |
Product
Corin Operating Tables Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; All Software Versions.
Reason
Under certain conditions, such as the simultaneous use of the operating table with other medical devices, including high-frequency (HF) surgical equipment, interference may occur that could prevent the device from performing as intended. If the issue occurs, the operating table is unresponsive to command from both the universal remote control and override panel.
Identifiers
| code_info | Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768…Model Numbers: (1) 770001A0, (2) 770001A2, (3) 770001B0, (4) 770001B2, (5) 770001F0, (6) 770001F2; UDI-DI: (1) 04046768145177, (2) 04046768145184, (3) 04046768145191, (4) 04046768145207, (5) 04046768145214, (6) 04046768145221; Lot/Serial Numbers: (1) 1, 2, 3, 10, 25, 26, 27, 28, 22, 4, 5, 6, 7, 8, 9, 11, 12, 13, 14, 15, 16, 17, 18, 19, 20, 21, 24, (2) 9, 51, 53, 14, 15, 16, 17, 34, 35, 36, 37, 39, 40, 42, 43, 54, 19, 20, 21, 22, 23, 24, 25, 26, 27, 28, 29, 30, 31, 32, 33, 38, 44, 45, 46, 5, 13, 57, 58, 59, 18, 49, 47, 48, 7, 52, (3) 41, 5, 6, 45, 48, 49, 50, 23, 38, 10, 12, 13, 14, 84, 87, 88, 89, 7, 63, 80, 82, 83, 86, 54, 93, 94, 11, 15, 18, 20, 21, 22, 27, 28, 32, 40, 9, 69, 70, 71, 72, 73, 74, 75, 76, 77, 78, 79, 85, 55, 56, 57, 58, 59, 42, 97, 92, 17, 26, 16, 24, 25, 30, 33, 35, 36, 37, 39, 44, 53, 91, 8, 19, 90, 46, 47, 51, 52, 34, 43, 60, 62, 64, 65, 66, 67, 68, (4) 44, 70, 71, 42, 56, 65, 115, 185, 52, 105, 106, 107, 109, 145, 146, 147, 148, 11, 26 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:57 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2420-2025%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.