FDA Device recall Z-2418-2018

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153V…

On 2018-07-18 the FDA classified a Class II recall of Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153V… by Medtronic Cardiac Rhythm And Heart Failure Crhf, citing medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2418-2018
ClassificationClass II
Statusterminated
Initiated2018-06-21
Published2018-07-18
Last updated2026-09-13 11:54 UTC
CompanyMedtronic Cardiac Rhythm And Heart Failure Crhf
DistributionUS

Product

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.

Reason

Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.

Identifiers

code_infoProduct UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.