FDA Device recall Z-2418-2018
Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153V…
- FDA Device
- Class II
- terminated
- Published 2018-07-18
On 2018-07-18 the FDA classified a Class II recall of Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153V… by Medtronic Cardiac Rhythm And Heart Failure Crhf, citing medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2418-2018 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2018-06-21 |
| Published | 2018-07-18 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Medtronic Cardiac Rhythm And Heart Failure Crhf |
| Distribution | US |
Product
Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.
Reason
Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.
Identifiers
| code_info | Product UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2418-2018%22