MINOR CATH LAB PACK DYNJ84775A OR VASCULAR ANGIO PACK DYNJ47269D PORT-A-CATH DYNJ63276G

On 2026-06-24 the FDA classified a Class II recall of MINOR CATH LAB PACK DYNJ84775A OR VASCULAR ANGIO PACK DYNJ47269D PORT-A-CATH DYNJ63276G by Medline Industries, citing kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2417-2026
ClassificationClass II
Statusongoing
Initiated2026-04-27
Published2026-06-24
Last updated2026-09-13 11:58 UTC
CompanyMedline Industries
DistributionUS

Product

MINOR CATH LAB PACK DYNJ84775A OR VASCULAR ANGIO PACK DYNJ47269D PORT-A-CATH DYNJ63276G

Reason

Kits contain recalled lot of Ferndale Laboratories MASTISOL Liquid Adhesive. Butyrate tubing may crack during actuation and render product unusable for application.

Identifiers

code_infoDYNJ84775A UDI-DI 10195327606527 lot 25LMH070 DYNJ47269D UDI-DI 10195327674786 lot 25LMJ234 DYNJ63276G UDI-DI 10198459458835 lot 26BBC598