RecallRadar

FDA Device recall Z-2417-2015

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873

On 2015-08-26 the FDA classified a Class II recall of Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873 by Radiometer America, citing when the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2417-2015
ClassificationClass II
Statusterminated
Initiated2015-07-16
Published2015-08-26
Last updated2026-09-13 11:53 UTC
CompanyRadiometer America
Distributionn/a

Product

Radiometer Fixation Kit N20 for the TCM CombiM monitoring system, REF 905-873. The Fixation Kit N20 is used to apply tc Sensor 84 and tc Sensor 54 on the skin of a patient. TC sensor 84 and 54 are used with the TCM CombiM monitoring system and the Philips TcG10 module. The TCM CombiM monitoring system is intended for continuous transcutaneous monitoring of carbon dioxide (tcpCO2) and oxygen (tcpO2) partial pressures.

Reason

When the fixation ring is taken off the paper liner there is a potential for fractions of the paper liner's surface coating to stick to the surface of the ring.

Identifiers

code_info
Model #: 905-873 Lots: 407221, 407231, 407241, 407251, 408041, 408191, 408201, 408211, 408231, 408232, 408261, 408…Model #: 905-873 Lots: 407221, 407231, 407241, 407251, 408041, 408191, 408201, 408211, 408231, 408232, 408261, 408271, 409021, 409101, 409231, 409291, 410021, 410141, 410281, 411171, 411201, 411261, 412021, 412111, 412161, 412201, 501081, 501131, 501191, 501311, 502031, 502101, 502131, 502171, 502231, 502261, 503091, 503231, 503311, 504031, 504081, 504111, 504271, 505041, 505061, 505141, 505181

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2417-2015%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.