RecallRadar

FDA Device recall Z-2415-2023

Snowden-Pencer MIS DIAMOND-TOUCH GRASPER MCKERNAN FUNDUS 5MM, 32CM, RATCHETED, REF SP90-7012

On 2023-08-23 the FDA classified a Class II recall of Snowden-Pencer MIS DIAMOND-TOUCH GRASPER MCKERNAN FUNDUS 5MM, 32CM, RATCHETED, REF SP90-7012 by Carefusion 2200, citing specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2415-2023
ClassificationClass II
Statusongoing
Initiated2023-07-13
Published2023-08-23
Last updated2026-09-13 11:56 UTC
CompanyCarefusion 2200
Distributionn/a

Product

Snowden-Pencer MIS DIAMOND-TOUCH GRASPER MCKERNAN FUNDUS 5MM, 32CM, RATCHETED, REF SP90-7012; endoscopic surgical device

Reason

Specific lots of the product have the potential to exhibit cracking or breaking of the rotation knob due to the incorrect use of an adhesive in the manufacturing process.

Identifiers

code_infoUDI/DI 10885403159671, Lot Number B23

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2415-2023%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.