RecallRadar

FDA Device recall Z-2413-2020

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J

On 2020-07-08 the FDA classified a Class I recall of Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J by Arrow International, citing A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration fail….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2413-2020
ClassificationClass I
Statusterminated
Initiated2020-05-20
Published2020-07-08
Last updated2026-09-13 11:55 UTC
CompanyArrow International
Distributionn/a

Product

Arrow AutoCAT2 Intra-Aortic Balloon Pump AUTOCAT2 WAVE JAPANESE, Product Code IAP-0500J

Reason

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Identifiers

code_info
Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120…Serial Number Ranges: AUTOCAT2 WAVE JAPANESE IAP-0500J 140866W through 140867W 150117W through 150118W 150120W 150201W 150305W 150307W through 150309W 150623W through 150624W 150713W through 150716W 151114W 151209W 151213W 151219W 160503W 160511W through 160515W 160517W 160635W through 160636W 160640W through 160641W 160731W through 160733W 171105W 171117W 171119W through 171120W 171201W 180120W 180301W

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2413-2020%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.