FDA Device recall Z-2399-2021
TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to…
- FDA Device
- Class II
- terminated
- Published 2021-09-08
On 2021-09-08 the FDA classified a Class II recall of TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to… by Depuy Orthopaedics, citing software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, cau….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2399-2021 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2021-08-10 |
| Published | 2021-09-08 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Depuy Orthopaedics |
| Distribution | CA, CO, FL, IL, IN, MI, MN, NY, TX, WA |
Product
TRUMATCH CT CUT GUIDE KIT R- TruMatch Patient Specific Instruments are intended to be used as patient-specific surgical instrumentation to assist in the positioning of a joint replacement component intra-operatively and in guiding the marking of bone before cutting Part Number: 420916
Reason
Software: A coding error associated with the Fast3D Segmentation software. During the scanning process the three images required of the hip, knee and ankle may not be aligned, causing the anatomic landmark locations to potentially result in limb malalignment.
Identifiers
| code_info | Lot number/ GTIN: 28226, 28135, 27612, 27544, 27431, 27407, 26970 OUS: 27476, 27779 GTIN: 10603295384861; |
|---|