FDA Device recall Z-2399-2016
Invacare¿ Bariatric Bed w/o mattress & rails, Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a…
- FDA Device
- Class II
- terminated
- Published 2016-08-17
On 2016-08-17 the FDA classified a Class II recall of Invacare¿ Bariatric Bed w/o mattress & rails, Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a… by Invacare, citing the firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2399-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-03-15 |
| Published | 2016-08-17 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Invacare |
| Distribution | US |
Product
Invacare¿ Bariatric Bed w/o mattress & rails, Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
Reason
The firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.
Identifiers
| code_info | Model #: BAR600IVC; Serial #s:15BF054037 15BF004216 15BF004217 15BF012364 15EF005762 15EF005763 15EF012120 15EF0…Model #: BAR600IVC; Serial #s:15BF054037 15BF004216 15BF004217 15BF012364 15EF005762 15EF005763 15EF012120 15EF012121 15GF006881 15BF048555 15DF039686 15DF039687 15LF016359 15LF016361 15HF010997 15HF025587 15HF025588 15IF014204 15IF014205 15IF014206 15IF014207 15IF014208 15IF014209 16AF014283 15DF013421 15DF027839 15DF027840 15DF027841 15DF027842 15EF025665 15EF025676 15HF002806 15HF002807 15HF002808 15HF002809 16BF020092 15IF040576 15JF003246 15KF016696 15KF016697 15LF001343 16AF030663 15EF027538 15EF027541 15JF035242 15JF035243 15JF035244 15JF035245 15JF035246 15LF032425 15LF032426 15GF003764 15JF040948 16AF016782 15DF019105 15DF033803 15DF033804 15EF005756 15EF031153 15EF031163 15EF031164 15EF031165 15DF001325 15EF025678 15FF022339 15FF037337 15HF025599 15IF024195 15IF036443 15LF009178 15LF039840 16AF041543 16AF041549 16BF002993 15LF014902 15FF022337 15BF042553 15BF042554 15CF023283 15CF023284 15DF022408 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2399-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.