FDA Device recall Z-2398-2024
Infusomat 60 DROP METRISET PUMP SET, 3 CARESITES Used with an electrically-powered infusion pump for intravenous (IV) administration of med…
- FDA Device
- Class II
- ongoing
- Published 2024-07-24
On 2024-07-24 the FDA classified a Class II recall of Infusomat 60 DROP METRISET PUMP SET, 3 CARESITES Used with an electrically-powered infusion pump for intravenous (IV) administration of med… by B Braun Medical, citing potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prim….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2398-2024 |
| Classification | Class II |
| Status | ongoing |
| Initiated | 2024-06-17 |
| Published | 2024-07-24 |
| Last updated | 2026-09-13 11:57 UTC |
| Company | B Braun Medical |
| Distribution | US |
Product
Infusomat 60 DROP METRISET PUMP SET, 3 CARESITES Used with an electrically-powered infusion pump for intravenous (IV) administration of medication Model/ Catalog Number: 490101
Reason
Potential for the backcheck valve to malfunction, resulting in backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime, resulting in patient injuries of adverse drug reaction and/or product malfunction with loss of medication/blood.
Identifiers
| code_info | UDI-DI 04046964186103; All Lots Distributed After 01AUG2023 |
|---|