FDA Device recall Z-2394-2016
ARTIS ZEE AND ZEEGO, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography sy…
- FDA Device
- Class II
- terminated
- Published 2016-08-17
On 2016-08-17 the FDA classified a Class II recall of ARTIS ZEE AND ZEEGO, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography sy… by Siemens Medical Solutions Usa, citing the possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2394-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-07-13 |
| Published | 2016-08-17 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Siemens Medical Solutions Usa |
| Distribution | US |
Product
ARTIS ZEE AND ZEEGO, system, x-ray, angiographic AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen are a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen family include cardiac angiography, neuro angiography, general angiography, rotational angiography, operating room angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures. AXIOM Artis, Artis zee / zeego and Artis Q/ Q.zen systems can also support the acquisition of position triggered imaging for spatial data synthesis.
Reason
The possibility exists for contaminants, in the form of biomass, to develop in the cooling system of Artis systems resulting in damage to the pump system.
Identifiers
| code_info | Model Numbers:10094135, 10094137, 10094139, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848…Model Numbers:10094135, 10094137, 10094139, 10094141, 10094142, 10094143, 10280959, 10848280, 10848281, 10848282, 10848283, 10848353, 10848354, 10848355, 5904433, 5904441, 5904466, 5904649, 5904656, 5917054, 7007755, 7008605, 7412807, 7413078, 7555357, 7555365, 7555373, 7727717, 7728350, 7728392 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:54 UTC. Current status: terminated.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2394-2016%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.