INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

On 2023-08-16 the FDA classified a Class II recall of INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE by Canon Medical System Usa, citing the C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2392-2023
ClassificationClass II
Statusongoing
Initiated2023-06-22
Published2023-08-16
Last updated2026-09-13 11:56 UTC
CompanyCanon Medical System Usa
DistributionUS

Product

INFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE

Reason

The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.

Identifiers

code_infoUDI-DI: 04987670100192. INFX-8000C, Serial Numbers: BHA1992001, BHB2132003, SWC2242002, AFB1832003, SUC2262008, SUC2292010, BFB1972004, SUB2172002, SWC2262003

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.