RecallRadar

FDA Device recall Z-2390-2020

Medtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler

On 2020-06-24 the FDA classified a Class II recall of Medtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler by Medtronic Sofamor Danek Usa, citing product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may incr….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2390-2020
ClassificationClass II
Statusterminated
Initiated2020-04-24
Published2020-06-24
Last updated2026-09-13 11:55 UTC
CompanyMedtronic Sofamor Danek Usa
DistributionUS

Product

Medtronic Bone Void Filler, Small - Product Usage: is an injectable, moldable and biocompatible calcium phosphate bone void filler. The single-use Medtronic Bone Void Filler Kit contains the necessary components for mixing of the bone void filler.

Reason

Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.

Identifiers

code_infoProduct number: C07A; GTIN: 00813845020504; Lot Number: 1127301

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:55 UTC. Current status: terminated.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2390-2020%22