FDA Device recall Z-2383-2012
Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators
- FDA Device
- Class II
- terminated
- Published 2012-09-26
On 2012-09-26 the FDA classified a Class II recall of Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators by Varian Medical Systems Oncology Systems, citing some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw w….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2383-2012 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2012-08-14 |
| Published | 2012-09-26 |
| Last updated | 2026-09-13 11:53 UTC |
| Company | Varian Medical Systems Oncology Systems |
| Distribution | n/a |
Product
Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Reason
Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
Identifiers
| code_info | LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,…LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H2 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:53 UTC. Current status: terminated.
Other recalls by Varian Medical Systems Oncology Systems
- Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Vari…
FDA Device · Class II · 2013-05-22 · Varian Medical Systems Oncology Systems - Varian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage: The RPM device is to be used…
FDA Device · Class II · 2013-05-01 · Varian Medical Systems Oncology Systems - Varian Eclipse Treatment Planning System for Radiotherapy
FDA Device · Class II · 2013-02-27 · Varian Medical Systems Oncology Systems - Varian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import Export Application, Model Number: HIT, Af…
FDA Device · Class II · 2012-12-26 · Varian Medical Systems Oncology Systems - ARIA Radiation Oncology, Version 11
FDA Device · Class II · 2012-11-07 · Varian Medical Systems Oncology Systems
All Varian Medical Systems Oncology Systems recalls
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2383-2012%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.