FDA Device recall Z-2380-2023
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637
- FDA Device
- Class I
- ongoing
- Published 2023-08-23
On 2023-08-23 the FDA classified a Class I recall of MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637 by Covidien, citing catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2380-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-06-28 |
| Published | 2023-08-23 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Covidien |
| Distribution | US |
Product
MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637
Reason
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Identifiers
| code_info | GTIN: 20884521057026 LOT Numbers: 2119400209 2122900117 2124600187 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2380-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.