FDA Device recall Z-2378-2023

MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629

On 2023-08-23 the FDA classified a Class I recall of MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629 by Covidien, citing catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2378-2023
ClassificationClass I
Statusongoing
Initiated2023-06-28
Published2023-08-23
Last updated2026-09-13 11:56 UTC
CompanyCovidien
DistributionUS

Product

MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 20 cm, Straight Extensions, Kit REF: 8888340629

Reason

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Identifiers

code_infoGTIN: 20884521057019 LOT Numbers: 1921300069 2227800100

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.