FDA Device recall Z-2377-2016
Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reu…
- FDA Device
- Class II
- terminated
- Published 2016-08-17
On 2016-08-17 the FDA classified a Class II recall of Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reu… by Stryker Howmedica Osteonics, citing stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2377-2016 |
| Classification | Class II |
| Status | terminated |
| Initiated | 2016-06-23 |
| Published | 2016-08-17 |
| Last updated | 2026-09-13 11:54 UTC |
| Company | Stryker Howmedica Osteonics |
| Distribution | US |
Product
Stryker Orthopaedics Modular Handle Triathlon Instruments, Non-Sterile, REF 6541-4-808, Lot 123, NDC M546654148080S Product Usage: Is a reusable instrument that is part of the Triathlon Primary total knee arthroplasty (TKA) system as well as the Triathlon Partial Knee Replacement (PKR) system. The use of the handles in Primary TKA is optional and may be used to: (1) Assist in stabilizing the Triathlon Femoral Sizer, (2) Aid in stabilization during use and removal of the Triathlon 4:1 Cutting Block and (3) Aid in assembly of the Anterior MIS Distal Resection Guide to the Anterior Skim Cut Guide. Per the Triathlon PKR surgical protocol, the handle is required for a PKR surgery and is used to: (1) Assemble and disassemble the Femoral Trial Drill Guide into the Femoral Trial and (2) Remove the Peg Trial from the Femoral Trial.
Reason
Stryker Orthopaedic is voluntarily recalling the Triathlon Modular Handle because analysis revealed that the press fit specifications between the dowel pin and the mating hole in the shaft were not being achieved potentially resulting in disassociation from the instrument.
Identifiers
| code_info | Catalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M…Catalog No. 6541-4-808 Lot Nos.: SB1K14, SB1K14A, SB1K14A1, SB1K14P, SB1K15, SB1K15A, SB1K15A1, SB1K15A2, SB1M25, SB1M25A, SB1M25A1, SB1M25A2, SB1M25A2X1, SB1N02, SB1N02A, SB1N02A1, SB1N02A2, SB1N02A3, SB1N02X1, SB1T05, SB1T05A, SB1T05A1, SB1T05AX1, SB1T05AX2, SB1T05KSB1V01, SB1V01K, SB1V01L, SB1V01X1, SB1V01X2, SB1V01X3, SB1W14, SB1W14K, SB1W14X1, SB2T19, SB2T19A, SB2T19D, SB2T19DX1, SB2T19DX2, SB2T19J, SB2T19K, SB2T19KX1, SB2T19KX2, SBT19X3, SB2V87, SB2V87A, SB2V87J, SB2V87JX1, SB2V87K, SB2V87X, SB2V87X1, SB2W23, SB2W23G, SB2W23X, SB2W23X1, SB3A20, SB3A20A, SB3A21, SB3A21D, SB3A21M, SB3A21X, SB3A22, SB3A22A1, SB3A22X1, SB3A55, SB3A55A, SB3A55A2, SB3A55AX1, SB3A55M, SB3A55X1, SB3H25, SB3H25D, SB3H25G, SB3H25X1, SB3H26, SB3H26D, SB3H26DX1, SB3H26X2, SB3K09, SB3K09X1, SB3K10, SB3K10L, SB3K10M, SB3K10MX1, SB3K10X1, SB3L19, SB3L19D, SB3L19K, SB3L19L, SB3L19P, SB3L19P1, SB3L20, SB3L20K, SB3L45, SB3L45A, SB3L45A1, SB3L45A2, SB3L45D, SB3L45DX1, SB3L45L, SB3L45T, SB3M17, SB3M17A, SB3M17K, SB |
|---|