FDA Device recall Z-2375-2026

Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2

On 2026-06-24 the FDA classified a Class I recall of Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2 by Arrow International, citing lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited rep….

Read the official FDA Device notice

AgencyFDA Device
Recall numberZ-2375-2026
ClassificationClass I
Statusongoing
Initiated2026-04-21
Published2026-06-24
Last updated2026-09-13 11:58 UTC
CompanyArrow International
DistributionUS

Product

Description/REF: PI CVC KIT: 3L 7 FR X 16 CM AGB+/ASK-42703-PNHS2; PI CVC KIT: 3L 7 FR X 20 CM AGB+/ASK-45703-PNHS2; PI CVC KIT: 4L 8.5 FR X 20 CM AGB+/ASK-45854-PNHS2

Reason

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Identifiers

code_info
REF(Material)/UDI-DI/Lot(Batch): ASK-42703-PNHS2/10801902193275/33F25J0107; ASK-45703-PNHS2/10801902193275/33F25B0593;…REF(Material)/UDI-DI/Lot(Batch): ASK-42703-PNHS2/10801902193275/33F25J0107; ASK-45703-PNHS2/10801902193275/33F25B0593; ASK-45854-PNHS2/10801902193275/33F24H0128

Status history

No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:58 UTC. Current status: ongoing.

Source

Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2375-2026%22

RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.