FDA Device recall Z-2374-2023
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, PASS Tray REF: 8888103002HP
- FDA Device
- Class I
- ongoing
- Published 2023-08-23
On 2023-08-23 the FDA classified a Class I recall of MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, PASS Tray REF: 8888103002HP by Covidien, citing catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter….
Read the official FDA Device notice
| Agency | FDA Device |
|---|---|
| Recall number | Z-2374-2023 |
| Classification | Class I |
| Status | ongoing |
| Initiated | 2023-06-28 |
| Published | 2023-08-23 |
| Last updated | 2026-09-13 11:56 UTC |
| Company | Covidien |
| Distribution | US |
Product
MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, PASS Tray REF: 8888103002HP
Reason
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
Identifiers
| code_info | GTIN: 20884521128023 LOT Numbers: 2133700232 2133700247 2133700248 |
|---|
Status history
No status changes recorded since RecallRadar first saw this recall on 2026-09-13 11:56 UTC. Current status: ongoing.
Other recalls by Covidien
- EndoFLIP System that includes: ENDOFLIP 1 SYSTEM EF-100 05391530810050 10884521809307 REFURBISHED SYSTEM RFG-EF100 ENDOFLIP 1 1088452180809…
FDA Device · Class II · 2026-08-26 · Covidien - V-Loc 90 Absorbable Wound Closure Device VLOC 90 0 VL 22 GS-21 VLOCM0336 VLOC 90 0 VL 22 GS-25 VLOCM0436 VLOCM0436 VLOCM0003 V-LOC* 90 4-0…
FDA Device · Class II · 2026-08-19 · Covidien - V-Loc 180 Absorbable Wound Closure Device BOX BEDSIDE STAPLING IS KING BOX06614V4 BOX BY PASS FINAL BOX07405V1 BOX BYPASS CARDON BOX01056V2…
FDA Device · Class II · 2026-08-19 · Covidien - Dermalon 88861757-31 DERMALON* 4-0 BLU 75CM C13 88861757-31 88861799-31 DERMALON* 4-0 BLU 75CM C14 88861799-31
FDA Device · Class II · 2026-08-19 · Covidien - Surgipro II S8698GX SURGIPRO II BL 5/0 18 P-13 S8698GX
FDA Device · Class II · 2026-08-19 · Covidien
Source
Official record: https://api.fda.gov/device/enforcement.json?search=recall_number:%22Z-2374-2023%22
RecallRadar indexes public recall notices from the FDA, USDA FSIS, and CPSC. The official agency notice is the authoritative source. Always confirm details at the linked source before acting.